Thermocouple vs RTD: choosing a temperature sensor for testing

Range, accuracy, cold junction compensation, 2/3/4-wire lead resistance and response time compared, with matching ICP DAS input modules.

Thermocouple vs RTD: choosing a temperature sensor for testing
Photo: Pete Wright · CC0 · Wikimedia Commons

Temperature is the most common measurement on a test bench, and the first decision is the sensor type. Thermocouple vs RTD is a choice between range and robustness on one side and accuracy and stability on the other. This article compares the two sensor types, explains the wiring details behind most measurement errors, and lists the ICP DAS modules that read each type.

How the two sensors work

A thermocouple is a junction of two different metals. It produces a small voltage that depends on the temperature difference between the measuring junction and the point where the thermocouple wires meet copper, the cold junction. A type K thermocouple gives roughly 41 µV per °C, so the input must resolve microvolts and reject noise.

An RTD (resistance temperature detector) is a resistor whose value rises with temperature. A Pt100 has 100 Ω at 0 °C and changes by about 0.385 Ω per °C; a Pt1000 has 1000 Ω at 0 °C. The input module drives a small current through the sensor and measures the voltage across it.

Thermocouple vs RTD at a glance

PropertyThermocoupleRTD (Pt100, Pt1000)
Temperature rangeWide: type T about −200 to +350 °C, types J and K up to about 750 °C and 1,200 °C, types R, S and B higher stillTypically −200 to +600 °C, some designs to +850 °C
Standard toleranceType K class 2: ±2.5 °C or 0.75 % of reading, whichever is largerClass B: ±0.3 °C at 0 °C, ±0.8 °C at 100 °C
Long-term stabilityCan drift, especially at high temperaturesVery stable
SignalMicrovolts; needs cold junction compensationResistance; needs excitation current and lead compensation
ResponseFast with thin or exposed junctionsSlower, depending on element and sheath size
CableThermocouple or matching extension cable to the inputCopper cable, ideally 3 or 4 wires

Choose an RTD when you need accuracy and stability between about −200 and +600 °C. Choose a thermocouple for higher temperatures, small or fast probes, and harsh mechanical conditions.

Cold junction compensation

A thermocouple measures a difference. To get the absolute temperature, the module measures the temperature at its own terminals and adds the equivalent thermocouple voltage. This cold junction compensation (CJC) is only as good as the temperature match between the terminals and the CJC sensor, so mount modules away from heat sources and draughts. Run thermocouple or matching extension cable all the way to the input terminals; a copper section in between creates new junctions at an unknown temperature.

Among ICP DAS inputs, the ET-2218HS-16 and USB-4018HS compensate automatically on every channel. The M-7018-16-G and the SG-3011 conditioner provide an adjustable CJC offset for fine calibration. For PCI-1800U boards, the DB-8225 terminal board has a cold junction circuit on analog input channel 1.

Lead resistance: 2-, 3- and 4-wire RTD connections

In a 2-wire connection, the lead resistance adds directly to the sensor resistance.

Worked example: a Pt100 on 10 m of 0.5 mm² copper cable sees about 0.34 Ω per conductor, or 0.69 Ω in the loop. Divided by 0.385 Ω/°C, the reading is about 1.8 °C too high. The same cable with a Pt1000 gives an error of about 0.18 °C, because the sensor changes ten times as much per degree.

A 3-wire connection uses a third lead to measure and subtract the lead resistance, assuming all leads are equal, and is the usual choice for test-bench wiring. A 4-wire connection carries the excitation current on one pair and measures the voltage on the other, which removes lead resistance from the result; use it for calibration and reference measurements.

The M-7015P-G, ET-2215H, ET-7015 and ET-7215 provide 3-wire lead resistance elimination. The M-7033-G accepts 2-, 3- and 4-wire RTDs, with lead resistance elimination for 3- and 4-wire connections.

Response time and sample rate

Sensor response time depends mostly on construction and thermal contact with the test object. Thin, exposed thermocouple junctions react fastest; a sheathed probe of either type reacts more slowly as its diameter grows.

The input module must also sample fast enough. Standard multi-channel modules share their sample rate across channels: the M-7018-G samples 10 Hz in total over 8 channels and the M-7015-G 12 Hz over 6, which suits slow thermal tests. For fast transients, the ET-2218HS-16 and USB-4018HS sample thermocouples at 100 Hz per channel, and the ET-2215H samples RTDs at 90 Hz per channel in fast mode, with ±0.1 % of FSR accuracy against ±0.05 % in normal mode at 1.5 Hz. The SG-3011H thermocouple conditioner has a switch-selectable response time of 0.5 ms or 100 ms.

Which ICP DAS modules fit

Thermocouple inputs

  • M-7018-G (8 channels) and M-7018-16-G (16 channels): RS-485 modules with Modbus RTU and DCON, differential inputs for types J, K, T, E, R, S, B, N and C, plus voltage and current ranges. The M-7018R-G adds 240 Vrms overvoltage protection and open thermocouple detection.
  • M-7019R-G: universal 8-channel input for thermocouples (including types L, M and L DIN43710), voltage and current, with an individual range per channel.
  • ET-2218HS-16: 16-channel Ethernet module with standard thermocouple connectors, 100 Hz per channel and automatic CJC.
  • USB-4018HS and USB-4018HS-16: 8 or 16 channels on USB, bus powered, ±0.5 °C accuracy and 100 Hz per channel.

RTD inputs

  • M-7015-G: 6-channel RS-485 module for Pt100, Pt1000, Ni120, Cu50, Cu100 and Cu1000 with open wire detection; the M-7015P-G adds 3-wire lead resistance elimination. The M-7033-G has 3 channels for 2-, 3- or 4-wire RTDs.
  • ET-2215H and ET-2215H-16: 8 or 16 high-speed RTD channels on Ethernet, 90 Hz per channel.
  • ET-7015 and ET-7215: 7-channel Ethernet RTD modules designed for long-distance RTD wiring.

Signal conditioners

For DAQ boards or PLCs with voltage or current inputs, the SG-3011 (thermocouple) and SG-3013 (RTD) convert one sensor to a voltage or current output, with 1000 VDC three-way isolation.

Send your sensor types, temperature ranges, channel count and required sample rate to our technical sales team, and we will suggest suitable modules.

Products for this application

PET-7017-10 CR
PET-7017-10 CR PET-7017-10 PoE Ethernet I/O Module, 10/20-ch Analog Input

The PET-7017-10 is a PoE-powered Ethernet analog input module for engineers who need voltage or current measurement on a networked I/O node. Its inputs can be set as 10 differential or 20 single-ended channels in software, and each channel can be given its own range, so one module can serve mixed sensor types on the same network.Each channel offers a choice of voltage ranges from ±150 mV to ±10 V and current ranges of ±20 mA, 0 ~ 20 mA and 4 ~ 20 mA, selected with jumpers. Resolution is 16-bit, and the module has two sampling modes: normal mode at 10 Hz total and fast mode at 50 Hz total. Open wire detection is available for 4 ~ 20 mA inputs, and differential channels are rated for 240 Vrms overvoltage protection.Configuration and monitoring are handled through a built-in web server with Web HMI, and the module supports Modbus TCP and Modbus UDP. Access is controlled by ID, password and IP filter. The unit is powered either from PoE (IEEE 802.3af, Class 1) or from a +12 ~ +48 VDC terminal block, and it mounts on a DIN rail.ApplicationsBuilding automationFactory automationMachine automationRemote maintenanceRemote diagnosisTesting equipment

Normaali hinta: 286,00 € per piece · volume discount from 10 pcs
DN-AIH-08 CR
DN-AIH-08 CR DN-AIH-08 HART Analog Input Termination Board, 8-ch Active/Passive Loop

DN-AIH-08 is a termination board for HART analog input modules. It measures the analogue signal on each of its 8 channels and can also supply isolated loop power to field devices, so it accommodates both active and passive loop wiring.The board includes short-circuit protection on the isolated power output. If the circuit is shorted or the current becomes too high, the loop current is limited. Removable fuses protect loop devices from high inrush current, and each fuse has a physical indicator that shows when it has blown.Suited to HART analog input installations where field devices need to be wired in either active or passive loop configuration.Mounts on DIN-rail or wall, with dimensions of 135 x 214.6 x 25.5 mm.ApplicationsConnecting HART analogue field devices to HART analog input modulesLoop-powered field instrumentation requiring isolated passive loop powerIndustrial installations requiring protected analogue loop wiring

Normaali hinta: 380,00 € per piece · volume discount from 10 pcs
DN-AII-08 CR
DN-AII-08 CR DN-AII-08 Current Input Termination Board, 8-ch Active/Passive Loop

DN-AII-08 CR is a termination board for current analog input modules. It connects field signals to the input channels and also measures the analog current, so it fits systems where the input is a current loop.The board supplies isolated loop power to field devices and supports both active and passive loop wiring on its 8 channels. The isolated supply is short-circuit protected: the loop current is limited when the circuit is shorted or the current becomes excessive. Removable fuses protect loop devices from high inrush current, and each fuse shows a physical indicator when it has blown.The board measures 135 x 214.6 x 25.5 mm and mounts on a DIN rail or on a wall. It is rated for an operating temperature of -25 ~ 75 °C and 5 ~ 95% RH, non-condensing.ApplicationsCurrent-loop field device wiring with active or passive loops

Normaali hinta: 329,00 € per piece · volume discount from 10 pcs
DN-AIO-M CR
DN-AIO-M CR DN-AIO-M Universal Termination Board for Analog I/O

The DN-AIO-M CR is a universal termination board for analog input and output modules. It is designed as a general-purpose wiring point for most kinds of analog module, so one board can serve a range of analog I/O installations.Its footprint is small, matching other mini termination boards, which makes it easier to fit into control panels where space is limited. It can be mounted on a DIN rail or on a wall.Pin assignment and wiring information covers voltage and current input and output configurations.Operating range of -25 ~ 75 °C and 5 ~ 95% RH, non-condensing.Carries CE, UKCA, FCC, RoHS and WEEE markings.ApplicationsTerminating field wiring for analog input modulesTerminating field wiring for analog output modulesCompact wiring points in control panels and installations

Normaali hinta: 191,00 € per piece · volume discount from 10 pcs

More on Test, measurement & data acquisition

Need help choosing?

We know the ICP DAS range and help you find the right module, accessory or replacement part.

Contact technical sales

For business customers

  • Quantity discounts applied automatically
  • VAT-free delivery within the EU with a verified VAT number
  • Datasheets and full specifications for every product
Create business account